List display of the measured values
KEW6315
- 93 -
KEW6315
*1
W screen: Displays of V and VL can be “customized” when “3P4W” is selected.
*2
W screen: “An” is displayed only when “3P4W” is selected.
*3
W screen: Displays of PA and C can be “customized” with the
(customize) Key. Line voltages
are converted into phase voltages to determine currents and phase angles for “PA” of 3P3W3A.
e.g.) Instantaneous values measured under 1P3W-2 (2 systems)
Symbol displayed on the LCD
V
*1
Phase voltage
VL
*1
Line voltage
A
Current
P
Active
power
+
consumption
Q
Reactive
power
+
Lagging
S
Apparent power
-
regenerating
-
leading
PF
Power
factor
+
Lagging
f
Frequency
-
leading
DC1
Analog input
Voltage at 1ch
DC2
Analog input
Voltage at 2ch
An*
2
Neutral current
PA
*3
V/A Phase
difference
+
Lagging
C
*
3
Capacitance calculation
-
leading