Martel Electronics LC100 User Manual

Page 12

Advertising
background image

D. Milliamp Measure with Loop Power

Again, Figure 5 shows the proper connections.
But in this case the device under test is a two-
wire transmitter that requires an external power
supply to generate an output current. In this
application the LC100 will supply 24 volts to the
loop while simultaneously measuring the result-
ant loop current. To turn on the loop power

supply press the

key while in the

mA measure mode.

Figure 5 - Measuring 4 to 20 mA

NOTE: The loop power mode can only be
enabled while in the mA measure mode.

E. Voltage Measure

Figure 6 shows how to connect the LC100 to
measure DC voltage up to 28 volts. To activate
the volts measure function simply press the

key to toggle through the functions

until the volts measure mode is reached. The

MODE

LOOP PWR

10

Advertising