Problem

Probable Cause

Try...

Check plumbing

Check the presence of Flow LED (see pg. 9)

Connect to earth ground

Check for proper electrical wiring

Check for power across power

input terminals

Note flow direction arrow,

reverse direction of meter

Reverse connections

Change output connections

Select another flow meter

Check for proper ground

Check for proper electrical wiring

Select another flow meter

Check for full pipe or install meter in the verti-

cal position

Check for ten diameters upstream AND five

diameters downstream

Install chemical injection line downstream of

magmeter (or far enough upstream to allow

complete mixing of fluids before meter)

Pipe not full

Below minimum flow cutoff

Unit not grounded

Excessive electrical noise

No power

Flow reversed

Power reversed

Output connections reversed

Fluid conductivity <20 microSiemens/cm

Missing or incorrect ground wire

Excessive electrical noise

Fluid conductivity <20 microSiemens/cm

Empty pipe

Not enough straight pipe

Rapidly changing conductivity (in chemical

injection or fertigation applications)

CAUTION & TROUBLESHOOTING

Seametrics Incorporated • 19026 72nd Avenue South • Kent, Washington 98032 • USA

(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • www.seametrics.com

TROUBLESHOOTING

Caution:

The electronics of the EX-Series meters are not field-repairable.

Warranty is void if unauthorized repair is attempted

No pulse output

Output pulses

incorrect

Jumpy reading