PIPING

Metallic pipe or tube, or plastic tubing can be used with the meter. The -025 and -038 meters have 3/8 NPT male

pipe fittings, and the -050 has 1/2 NPT male pipe fittings. For plastic tubing, use female NPT x tubing adaptors. In

installations exposed to vibration, using the piping as the sole support for the meter is not recommended, particularly

on the smaller sizes.

If metal piping is used, ground connections from the grounding studs on the bottom of the unit to both inlet and outlet

pipes provide the best results in metering accuracy. This does not apply to plastic tubing. A minimum of 1” of straight

pipe run is recommended on the inlet side of the meter. Straight run is not necessary on the outlet.

CONNECTIONS

Power and signal connections are provided by the single 4 or 6-conductor cable, and are color-coded. See the “wiring

connections” chart. In addition, it is necessary for proper operation to ground the unit to a good quality earth ground.

Use the 12 ft. grounding wire supplied with the unit, or if necessary, a longer wire attached to one or both of the

grounding studs.

WIRING
CONNECTIONS

EM 100/102 SERIES

MAGNETIC FLOW METER INSTRUCTIONS

RED + 12 to 24 VDC Power

BLACK - Power

GREEN + Pulse Output Isolated

WHITE - Pulse Output Isolated

ORANGE + 4-20 mA Isolated

BLUE - 4-20 mA Isolated

}

Optional