Connections, Operation – Seametrics FT520 v.2 User Manual

Page 3

Advertising
background image

3 of 4

Connections

Set Decimal. Use

to select zero, one, or two decimal

places on the flow units. Press

to advance to the

next menu.

Set K-Factor. The unit will not function properly until
this number is entered. The K-factor is simply the number
of pulses which the flow meter or flow sensor puts out
per gallon of liquid. It is marked on the Model/Serial tag
of SeaMetrics flow meters and flow sensor fittings. On
adjustable flow sensors, the K-factor must be taken from
the chart in the flow sensor instructions, based on pipe

size. Set the number with

keys. The digit which

is underlined is the one being set. The

moves one

digit to the left. The

increases the digit. When the

appropriate number has been set, press

again to

move on to the next menu item.

Set Pulse Out. An output pulse is activated at the
selected volume intervals if this feature is in use. If a
pulse output is not needed, pulse out does not need to

be set. Press

to move on to the next menu item.

Set 20 mA Out. The “SET 20mA RATE” programs the
flow rate at which the output is 20 mA, 5 V or 10 V. Use

and

to set the maximum flow rate for full scale

output. Press

for next menu item.

Clear Total. This function resets the running total back
to zero. It is unrelated to the regeneration function, and
can be ignored unless it is needed. To reset to zero,
press

and

simultaneously. Press

to return

to batch mode.

Operation

Start Batch. Pressing the Start/Stop key starts the batch
by energizing Relay 1. The indicator for Relay 1 will light,
indicating normal operation. The indicator for Relay 2
may also light, depending on the prewarn setting. If
staged shut-off is being used, the Relay 2 indicator light
should go out before the end of the batch, at the prewarn
set point.

The batch will continue on to the set amount unless it is
halted in the middle by pressing the Start/Stop key again.

Pulse 1

Alternate

12 VDC power

Pulse 2

Relay 2

Relay 1

NC

COM

NO

NC

NO

COM

Batch Outputs

Line

Neutral

Ground

Resume

0-5 V or
0-10 V or
4-20 mA

Sensor Input

Start/Stop

+

+

s1

+

s2

RESUME

COM

BATCH

Remote
Controls

Pulse Outputs

(for metering pumps)

Main

power fuse

250mA

part # 26926

AC Power

Ribbon connector

from display board

Meter

Power signal

RED

BLACK

Dry contact

Customer Supplies
1A Fuse

Ground

A
B

An Output

A

P

Passive

Active

Jumpers

5V

10V

4-20mA

PC1 power supply

P/N 05040

WHITE

Analog Switch

Analog Header