6comparator function – SHIMPO TRC-Series User Manual
Page 15

11
6
Comparator Function
•
This function compares the High/Low limits and measured data (Mean Mode) or peak data (Peak
Mode) or memory data. The result is displayed by Comparator LED indicators.
•
If both High and Low limits are set to 0, the Comparator function is deactivated. Refer to 4.2.2
Operation.
•
The buzzer sounds and the “Go” LED illuminates when the displayed data is between the High and
Low limits.
- In Peak mode, the buzzer sounds when the peak value is fixed.
- In Mean mode, the buzzer sounds every 1 sec.
Note: When the displayed value is less than 1% of the torque rating, the buzzer does not activate.
•
Comparison is used with all completed data.
<Example of comparator function>
High limit = 100, Low limit = 50
Measurement of CW direction
Measurement of CCW direction
-
100
High limit
50
Low limit
Hi-NG: ON
L0-NG: ON
+
0
Go: ON
-
50
Low limit
100
High limit
Lo-NG: ON
Hi-NG: ON
+
0
Go: ON