To connect ph, orp or ise electrode, To connect the temperature probe – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 10

Advertising
background image

6

3.2

Connecting the pH Electrode, Temperature Probe & Electrode Holder

The Eutech Instruments CyberScan pH 10/100 meter uses any standard pH, ORP, or Ion

Selective Electrode (ISE) with a BNC connector. For Automatic Temperature Compensation

(ATC), this meter requires a temperature probe with a phono-jack connector.

NOTE: It is important that water does not get onto the BNC connector. Also avoid touching the

connector with soiled hands.

3.2.1

To connect pH, ORP or ISE electrode

1.

Slide the electrode connector of the

electrode over the BNC connector

socket on the meter.

2.

Make sure the slots of the connector are

in line with the posts of the socket.

3.

Slide the BNC connector of the probe

over (Figure 6).

4.

Rotate and push connector clockwise

until it locks.

5.

To remove electrode, push and rotate

the connector anti-clockwise.

6.

While holding onto the metal part of

the connector, pull it away from the meter.

7.

Be careful not to use excessive force.

CAUTION: Do not pull on the probe cord or the probe wires might disconnect.

Refer to “Accessories” section for information on temperature probe and other electrodes.

3.2.2

To connect the temperature probe:

The temperature probe (provided) uses a phono

jack to connect with the socket on the CyberScan.

Insert the jack fully into the socket (Figure 7).

Figure 6: Connect pH or ORP electrode using BNC

Figure 7: Connecting the temperature

probe