32

9.3

Program 3 –Meter Configuration

The following three options are available to the user for

customization:

9.3.1

P3.0 : Resolution

The resolution of the instrument in the pH measurement

mode can be user selected to 0.1 or 0.01 pH. Under

default condition, the instrument is set for 0.01 pH

resolution.

LCD Value

Description

0.01

A resolution of 0.01 pH is shown.

0.1

A resolution of 0.1 pH is shown.

1. Use

MI/

!

or MR/

"

keys to select the resolution

desired. See Figure 33.

2.

Press CON key to continue to P3.1.

9.3.2

P3.1 : Ready selection

Activation of the READY option ensures that the READY

indicator is displayed when the electrode reading has

stabilized.

The stabilization tolerances for the various measurement

ranges are as indicated below. Under default conditions,

i.e. when the [SETUP] mode is not programmed, the

instrument has this option activated.

MODE

VARIATION

RANGE

pH

± 0.01 pH

mV

± 0.8 pH

< 400 mV

mV

± 1.2 mV

> 400 mV

ON

READY Indicator will appear

when the reading stabilizes

within ± 0.01 pH

OFF

READY indicator will not appear.

Figure 33: Setting pH resolution

Figure 34: READY Selection