Troubleshooting hints – Thermo Fisher Scientific Ion Selective Electrodes Chloride User Manual

Page 19

Advertising
background image

Instruction Manual

Chloride Electrode

TROUBLESHOOTING HINTS

Symptom Possible Causes

Next Step

Out of Range defective meter

check meter with shorting

Reading

strap (see meter

instruction manual)


defective electrode

check electrode operation


electrodes not

plug electrode into position

plugged

in

properly

and

reset

meter

reference electrode

clean junction

junction clogged


static

electricity wipe

plastic

meter

face

with

detergent solution


meter

or

stirrer ground

meter

or

stirrer

not

grounded


Noisy or Unstable defective meter

check meter with

Readings (readings

shorting strap

continuously or
rapidly changing) air bubble on membrane

remove bubble by

re-dipping electrode


ISA not used

use recommended ISA


meter or stirrer

ground meter or

not grounded

stirrer


defective electrode

replace electrode


electrode exposed

soak electrode in

to interferences

chloride standard


Drift (reading

samples and standards

allow solutions to come to

slowly changing at different temperatures

room temperature before

in one direction)

measurement


incorrect reference

use recommended

filling solution

filling solution in outer

of reference electrode

membrane

dirty

or

etched

polish

membrane

19