Instruction Manual

Chloride Electrode

Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


ISA not used

use recommended ISA


standard used as ISA

use ISA



"Incorrect Answer" incorrect scaling

plot millivolts on the linear

(but calibration

of semi-log paper

axis. On the log axis, be

curve

is

good)

sure

concentration

numbers

within

each

decade

are

increasing with increasing

concentration


incorrect sign

be sure to note sign of

millivolt number correctly



incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

1.0x10

-3

M = 35.5 ppm as Cl-


complexing agents

remove complexing agent