A t c – Thermo Fisher Scientific Alpha CON 100 User Manual

Page 15

Advertising
background image

12

5.3 Calibrating the Controller for Temperature

Measurement


5.3.1 The

CAL

°

C menu

1. Enter this menu by pressing

MODE key twice to the “SEt”
display if you are in
measurement mode.

2. Press ENTER key. If the

password has been set earlier,
key in the password using the
method described in section 4.1.
Press ENTER once to confirm.
After you set the correct
password, you see the “CAL
and “COn” display blinks
alternatively. Press ▼ key once,
the display shows “CAL”

o

C”

blinking.

1.

Press ENTER and the display

alternatively shows “Atc” and “On” or “Atc” and “OFF”.
Use ▲ and ▼ keys to choose between both ATC ON and

ATC OFF.

ATC ON : If a PT 100 is connected,
use ▲ and ▼ keys to adjust the
temperature offset of the PT 100 by

±

5

°

C. Dip the probe into the sample

liquid. Make sure that the display is
set to ATC ON, press ENTER and the
display now shows the actual
temperature reading (blinking).

Use ▲ and ▼ keys to adjust the
reading to its actual temperature - as
measured by an external
thermometer. Once done, press
ENTER key and the display will flash
“CAL” “Con” alternatively.

SETUP

MODE

ENTER

SEt

SETUP

MODE

ENTER

CAL

SETUP

MODE

ENTER

o

C

SETUP

MODE

ENTER

A t c

SETUP

MODE

ENTER

On

SETUP

MODE

ENTER

23.5

Advertising