Thermo Fisher Scientific Alpha CON 100 User Manual

Page 20

Advertising
background image

17

5.4.4 The Lo HYS Menu
1. Enter this menu by pressing

the MODE key to the “SEt”
display if you are in the
measurement mode.

2. Press ENTER key. If the

password has been set
earlier, key in the password
using the method described
in section 4.1.

3. Press ENTER once to

confirm. After you set the
correct password, you see
the “CAL” and “COn
display blinks alternatively.
Press

key five times, the

display shows “Lo” “HYS”
blinking alternately.

4. Press

ENTER key to access

the Lo HYS menu and the
display shows the last Lo
HYS value or default (2% of
Full Scale). Use

and

keys to change the value of
the Lo HYS point.

5. Press ENTER to confirm the value of the Lo HYS.

Press MODE key to exit to the measurement mode.
You can press MODE key (as an ESCAPE key) to
revert to the measurement value if ENTER is not
pressed; the set value is not stored into memory.


IMPORTANT : The Lo SET + ½ Lo HYS should be less
than Hi SET - ½ Hi HYS i.e. Hi SET and Lo SET value (in
consideration of the hysteresis band) can never overlap. Lo
HYS and Hi HYS can be set independent of each other; this
allows non-symmetrical hysteresis option.

NOTE : The maximum
value of Hysteresis is
4% of its Full Scale (FS)
selected. The Lo HYS
hysteresis is spread
equally on either side of
the Lo SET point.

SETUP

MODE

ENTER

Lo

SETUP

MODE

ENTER

02

SETUP

MODE

ENTER

HYS