General descriptions, 1006 dc millivolt source, Descriptions – Time Electronics 1007 DC Millivolt Potentiometer & Source User Manual

Page 3

Advertising
background image

3

1006/1007 Technical Manual

3

1. Descriptions

1.1. 1006 DC Millivolt Source

• 3 Ranges

0-1V (0.1 mV steps)

0-100mV (10 uV steps)

0-10mV (1 uV steps)

• 0.02% Accuracy

• 20mA Output Current

• Short Circuit & Overload Protected

• Battery Powered – Portable

• Safety Terminals

• Battery Level Indicator

The 1006 is an accurate low cost millivolt source suitable for voltage injection applications.

Three output ranges are provided to give adjustable output values from 1µV to 1V with a
basic 0.02% accuracy. For signal injection, the operator needs to switch on, check the
battery condition, select the range, and set the required voltage using the thumbwheel
switches. The 1006 uses a precision reference diode and low temperature coefficient
resistors to give a highly stable output.

Power is provided by 6 AA batteries. Battery life is several months, depending on usage.
The battery condition is monitored by an indicator mounted on the top of the unit. The 1006
has up to 20mA drive current and is short circuit and overload protected.
An off/normal/reverse output polarity switch is provided.

Operation: The 1006 is simple to operate and does not require standardisation or
calibration prior to use. For signal injection, the operator needs only to switch on, check the
battery condition and select the required range output.

Advertising
This manual is related to the following products: