Output: 0.999.9mV in 3 ranges

0.999.9mV in 0.1mV steps

0.99.99mV in 10uV steps

0.0.999mV in 1uV steps


Accuracy:

± 0.02% of setting

± 0.02% of range


Output Resistance:

Less than 0.1 ohm on 1V and 100mV ranges.

1 ohm on 10 mV range.


Maximum Output Current: 1V and 100mV ranges - 20mA.

10mV range: Up to short circuit value although it should be

noted that loads of less than 1K ohm will give greater than

0.1% error.


Output Voltage Stability:

Less than 60ppm/°C.

Less than 100ppm per 3 month. (non cumulative)


Operating Temperature:

-10°C to + 60°C.


Output Polarity:

Positive or negative switch selected. A centre ‘off’ position is

also provided.


Output Noise Level:

Less than 30ppm of f.s.


Reference Source:

Precision zenor diode selected for stability and low

temperature coefficient.


Maximum Overload:

The instrument can withstand continuous short circuit on the

output for all ranges.


Power Supply:

6 AA batteries. A battery condition display indicates when the
batteries should be changed. An alternative power source is 6
NiMH cells of the same dimensions - these can be recharged
via a socket on the top of the unit. The 6 rechargeable
batteries and mains re-charger unit are available as an
optional extra.


Null Balance Display:

The null display is on a front panel meter, zero and sensitivity

(1007 only)

controls are provided.


Maximum sensitivity: ± 20uV fsd (2uV/div.)

Minimum sensitivity: ± 20-0-20(20 divisions).

Input resistance: Greater than 1M ohm at balance.


Dimensions:

200 x 75 x 110mm (215 x 100 x 120mm incl. protective boot)


Weight:

1.1kg (1.5kg incl. protective boot)