M-Series® M5000

Firmware Upgrade

MAG-UM-00945-EN-02 (August 2014)

Installation Manual

PREPARING FOR THE FIRMWARE UPGRADE

Meters that currently have version 1.x firmware can be upgraded

only to the latest 1.x firmware. Meters that currently have version

2.x firmware can be upgraded to the latest 2.x firmware. If you have

1.x firmware and wish to upgrade to 2.x firmware, you will need to

upgrade the board.

Download the latest firmware version from our website at

www.badgermeter.com

to your hard drive. With the Flow Meter Tool

software installed, connect the laptop to the meter using either the

supplied RS232 cable or an IrDA adapter.

OTEE:

N

The M5000 Data Logging kit (p/n 67354-008) is needed to

perform the firmware upgrade. See the M5000 Data

Logging Installation Manual.

RS232 Connection

1. Identify / Configure the meter’s communication settings:

a. Navigate to Communications > Interface.

b. Set the interface to SERIAL.

OTEE:

N

The interface must be set to SERIAL. All other settings

can be set as desired by the operator and must match

those settings of the software tool.

c. Record or change other interface parameters (parity

and baud rate).

2. Connect the supplied RS232 cable into the RS232 connector of

the meter. Either connect the serial connector to a COM port or

connect it to the USB adapter.

3. Open the Flow Meter Tool installed on the laptop or PC. Go to

Start > All Programs > Badger Meter to open the flow meter tool

application.

4. To configure the flow meter tool software communication

settings:

a. Select Communication Settings.

b. Change the following parameters as necessary to

align with the meter settings:

MODBUS ADDRESS (Node Address)

BAUD RATE (9600)

DATA BITS (default is 8)

PARITY

STOP BITS (default is 1)

c. Select OK to confirm the configuration of the

communication port. Make sure you select the

correct COM port.

OTEE:

N

The M5000 communicates via the COM port and IrDA.

TCP/IP is not supported.

IrDA Connection

1. Identify / Configure the meter’s communication settings:

a. Navigate to Communications > Interface.

b. Set the interface to IrDA.

OTEE:

N

The interface must be set to IrDA. All other settings can

be set as desired by the operator and must match those

settings of the software tool.

3. On the M5000 enclosure, install the IrDA bracket. Then install

the IrDA cable.

4. Connect the IrDA cable from the M5000 to the laptop.

5. If necessary, install the rDA link drivers from the CD. The IrDA

link uses a USB/IrDA converter and drivers. You may have to

restart your computer after installing the drivers.