Insert suction fitting into pump head. Remove your
fingers from pump head. Start pump by pressing
START button. Grab hold of Tube Installation Tool
and use it to leverage tubing into pump head.
Introduce tubing into pump head while the rotor is
rotating. Avoid using fingers to guide the tubing.
Stop pump at anytime by pressing STOP button.
Start pump by pressing START button.
Continue to follow rotation of rotor while directing
tube into pump head. At this point, you may need to
pull Tube Installation Tool to stretch tubing into
position. Let rotor spin a few rotations while pulling
Installation tool so fitting can be properly installed.
Continue to pull Tube Installation Tool to allow
enough room to slide discharge fitting into pump
head tongue and groove. Once discharge fitting is
secured in pump head, stop pump by pressing STOP
button. Replace pump head cover. Pump will ask you
if you’d like to reset tube timer. If you choose yes,
current tube time will display for 5 seconds before
resetting to zero. Make note of your displayed tube
life. Select Yes again to reset tube life timer.
Re-attach Pump Head Cover using the four Thumb Screws.
Pump will detect Pump Head Cover is installed and begin to exit MAINTENANCE
MODE.
Pump will ask you if Tube was replaced. Yes / No
If Yes is selected, pump ask you to reset Tube Timer. Yes / No
If Yes is selected, pump will display Current Tube Timer briefly (5 seconds) before
resetting to zero.
The pump can now begin normal operation.
MAINTENANCE MODE
Pump Cover Detected!
Press ENTER to set
Was Tube Replaced?
MAINTENANCE MODE
Up/Down to select
YES
NO