Crystal MultiCal Pressure Module User Manual
Page 7

Calibration
5
MultiCal H2O/PSI Pressure Module Manual
Fine Zero
inH
2
O
psi
-
Linearity
R1
R12
R15
Coarse Zero
Potentiometer locations.
6
Check the inch of water pressure points shown in the following table. Adjust R12 if necessary to bring the readings to within the tolerances shown in
the table.
Readings in mV @ 23°C ±5°C
True Pressure
Acceptable Output (mV)
100 inH
2
O
99.85 to 100.15
415 inH
2
O
414.70 to 415.31
830 inH
2
O
829.49 to 830.52
30 psi
29.98 to 30.03
7
Set the switch on the MultiCal to the PSI position.
8
Test the MultiCal at the 30 psi point. The reading should be within the tolerance shown in the table above. If not, adjust R15 until both points are within
specification.
9
In the unlikely event that you are unable to adjust the MultiCal so that all of the points meet the tolerances of the table above, contact Customer Service for
assistance. Do not readjust linearity. The linearity adjustment is factory set and should never require readjustment, unless the sensor is replaced.