Temperature derating, General specifications – Crystal MultiCal Pressure Module User Manual
Page 9

Specifications
7
MultiCal H2O/PSI Pressure Module Manual
—
TEMPERATURE DERATING
Add to Basic Accuracy Specification. °C = ambient temperature.
Temperature Range
Derating
28 to 55°C . . . . . . . . . . . . . . . . . . . . . . . (0.016%/°C) x (°C – 28°C)
18 to 28°C . . . . . . . . . . . . . . . . . . . . . . . No Derating
0 to 18°C . . . . . . . . . . . . . . . . . . . . . . . . (0.048%/°C) x (18° – °C)
-10 to 0°C . . . . . . . . . . . . . . . . . . . . . . . . (0.264%/°C) x (9° – °C)
—
GENERAL SPECIFICATIONS
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . 164 grams (5.8 oz.) w/battery
Overall Length . . . . . . . . . . . . . . . . . . . 118mm (4 5/8")
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . NEDA # 1604, 6F22, 006P
Battery Life . . . . . . . . . . . . . . . . . . . . . . 400 Hours typical with alkaline battery
Output . . . . . . . . . . . . . . . . . . . . . . . . . . 1mV/unit into 10MΩ load
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . -51 to 71°C