Pressure ranges, display scales, and resolution – Crystal XP2i-DP Digital Differential Pressure Gauge User Manual
Page 7

Specifications
5
XP2i-DP Operation Manual
—
pressure ranges, dIsplay scales, and resolutIon
psi
bar
kPa
MPa
Overpressure (+/– ports)
psi
inH
2
0
inHg
mmHg
mmH
2
0
kg/cm
2
bar
mbar
kpa
Mpa
15PSI
1BAR
100KPA
0.1MPA
6.5x / 3.0 x
0.001
0.01
0.001
0.01
1
0.0001
0.0001
0.1
0.01
0.00001
100PSI
7BAR
700KPA
0.7MPA
2.0 x / 2.0 x
0.01
0.1
0.01
0.1
1
0.0001
0.0001
0.1
0.01
0.00001
•
Unneeded pressure units may be disabled via the RS-232 connector using ConfigXP software.
•
An XP2i-DP will indicate pressure up to 10% above its stated pressure rating. Above 110%, the XP2i-DP display will flash, indicating that the applied pressure
exceeds the calibrated pressure range. If the calibrated pressure range is exceeded, the pressure displayed may not be accurate.
•
Overpressure is the maximum pressure the gauge can withstand without damage. The gauge will not indicate pressure up to this value.
•
kPa and MPa models can display pressure in kPa, MPa, and bar (and/or mbar) only. psi and bar models can display all available units.