Specifications

5

XP2i-DP Operation Manual

pressure ranges, dIsplay scales, and resolutIon

psi

bar

kPa

MPa

Overpressure (+/– ports)

psi

inH

2

0

inHg

mmHg

mmH

2

0

kg/cm

2

bar

mbar

kpa

Mpa

15PSI

1BAR 

100KPA

0.1MPA 

6.5x / 3.0 x

0.001

0.01

0.001

0.01

1

0.0001

0.0001

0.1

0.01

0.00001

100PSI

 7BAR

700KPA

0.7MPA 

2.0 x / 2.0 x

0.01

0.1

0.01

0.1

1

0.0001

0.0001

0.1

0.01

0.00001

Unneeded pressure units may be disabled via the RS-232 connector using ConfigXP software.

An XP2i-DP will indicate pressure up to 10% above its stated pressure rating. Above 110%, the XP2i-DP display will flash, indicating that the applied pressure
exceeds the calibrated pressure range. If the calibrated pressure range is exceeded, the pressure displayed may not be accurate.

Overpressure is the maximum pressure the gauge can withstand without damage. The gauge will not indicate pressure up to this value.

kPa and MPa models can display pressure in kPa, MPa, and bar (and/or mbar) only. psi and bar models can display all available units.