Applications

dIfferentIal pressure

The XP2i-DP may be used to determine accurate differential pressure measurements across a range of applications or installations such as filters,
orifice plates, valves, etc. For example, in a piping system with an upstream static pressure of approximately 80 psig, the XP2i-DP may be used to
determine the 10 psid pressure drop across the device under test (DUT). In this case, the product would be connected as shown below.

FLOW

80

PSI

70

PSI

XP2i-DP

Device Under Test (DUT)

100 PSID

XP2i

Digital Test Gauge

Hi

Lo

peak

clear

zero

units

The pressure measurement accuracy of the gauge in this example is a combination of the standard accuracy statement and the static line pres-
sure effect. The 10 psid pressure, when compared to the 100 psi full scale product used, will require the 0 to 20% of Full Scale, ± (0.02% of Full
Scale) accuracy statement. The static line pressure effect relates to the lowest test pressure applied in the test, or in this case 70 psi.
The pressure measurement accuracy for this test condition is as follows:

± (0.02% x 100 psi) + (0.0003 psi x 70 psi) = ± 0.04 psi.

When using a 15 psi full scale gauge, this will require the 20 to 100% of Full Scale, ± (0.1% of Reading) accuracy statement.

± (0.1% x 10 psi) + (0.0003 psi x 70 psi) = ± 0.031 psi.