Pressure is our business – Crystal MultiCal BAR/PSI User Manual

Page 4

Advertising
background image

PN: 1763 Rev A, 4/2004
Page 4 of 7

PRESSURE

is Our BUSINESS

Temperature derating:

Range

Humidity

28 to 55 °C

0.016% x ( °C -28°)

18to 28 °C

No Derating

0 to 18 °C

0.048% x (18° - °C)

-10 to 0 °C

0.264% x (9° - °C)

(Add to Basic Accuracy Specification, °C = ambient temperature)

Wetted Materials:

316 stainless steel

General

Weight

230 grams (8.1 oz) w/battery

Overall length

118 mm (4 5/8”)

Pressure Fitting

1/8” NPT

Battery

NEDA #1604, 6F22, 006P

Battery Life

400 Hours typ. w/alkaline battery

Output

1mV/unit into 10 Mohm load

Storage

-51 to 71 °C

MultiCal Calibration:

A calibration cycle of 1 year is recommended to maintain the MultiCal within specifications.
Recommended equipment is listed in Table 1.

Table 1 : Recommended Equipment

Instrument

Minimum Specification

Recommended Model

Pressure Standard

± 0.025% of Reading

Pressurements

Dead weight gauge

Digital Voltmeter

± 0.025% of reading,
10 Mohm input impedance

Fluke 45

You do not need to use a reference multimeter if the MultiCal will be used with only one
multimeter. In those cases, your overall accuracy may be better if it is calibrated as a set with
the matching multimeter.
The recommended Digital Voltmeter listed in Table 1 has 10MΩ input impedance - as
do many handheld multimeters. Your reference multimeter should have the same input
impedance as the multimeters the module will be used with. In some cases this may mean
that the reference multimeter will need to have its input impedance lowered.
For example, a Flike 8840A is an excellent reference meter but has an input impedance Ω
10000MΩ. If the meters the module will be used with are Fluke 87s, connect a 10MΩ resistor
in parallel with the input to the 8840A.