PRESSURE

is Our BUSINESS

9. In the unlikely event that you are unable to adjust the MultiCal so that all of the points
meet the tolerances of Table 2, contact Customer Service for assistance. Do not readjust
linearity. The linearity adjustment is factory set and should never require readjustment,
unless the sensor is replaced.

Table 2: Readings in mV

True Pressure

Acceptable Output (mV)

1500 PSI

1498.3 to 1501.7

3000 PSI

2996.8 to 3003.2

206 PSI

205.774 to 206.226