9.1. RECEIVED SIGNAL
Select Tab Page 2 on the MAINTE screen and the following screen appears.
Fig. 9-2 RECEIVED SIGNAL screen
x Select one from forward direction received wave, reverse direction received wave, forward direction filter, reverse
direction filter and calculation waveform.
x Click the [START] button to read in the specified cycle and click the [Stop] button to cancel the operation. Display is
updated with each read.
x Save the received waveform in the specified file by [SAVE As CSV] button. Trigger level is also displayed.
x Display the RAS information in the STATUS DISPLAY.
1. Startup is completed within 3 to 6 waves.
2. No peak (amplitude) fluctuation should be observed.
Otherwise air is mixed in.
Received
signal
Trigger level
Point
Bulletin F