Dwyer instruments, inc – Dwyer WT2-10 User Manual

Page 2

Advertising
background image

4. Temperature Unit Selection (°C/°F )
With the meter on, press and hold the MODE button for more than 2 seconds to select
the desired units; the display will indicate the changes (°C/°F).

5. Auto Power OFF
The meter will shut off automatically after approximately 10 minutes if no action is
taken. To disable auto power off, when meter is off press HOLD button and extend the
probe away from the body of the meter, “n” icon will appear on the LCD, and release
HOLD button. The meter will return to normal measurement and “TEMP” icon will
appear with the temperature reading.

6. Setting Mode
Hi Lo Alarm
Step1. Hi alarm

7. Stop Beeper Temporarily
If the temperature exceeds the Hi/Lo alarm setting, the meter will give an audible
sound, shortly press MODE button to stop the beep sound, then the meter is ready to
take the next measurement.

8. Battery Replacement
Use a coin to remove the battery compartment cover on the rear of the meter. Replace
the CR-2032 lithium battery; install the new battery face up (+) in the battery
compartment.

Disposal: Follow the valid legal stipulations in respect of the disposal of the device at
the end of its life cycle.

MAINTENANCE/REPAIR
Upon final installation of the WT2-10, no routine maintenance is required. The Series
WT2-10 is not field serviceable and should be returned if repair is needed. Field repair
should not be attempted and may void warranty.

WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping the
product back for repair. Be sure to include a brief description of the problem plus any
additional application notes.

With meter on, press MX/MI and HOLD buttons at the same
time to enter Hi alarm setting. A blinking “A-on” icon will
appear on the LCD.
Press MODE button to select High alarm “A-on” or “A-oF”.
(a) If “A-on” is selected, press MX/MI button to save and
move to step 2 for High temp. alarm setting.
(b) If “A-oF” is selected, long press MX/MI button to save.

After Hi alarm is set, H150°C will appear on the LCD and
digit “1” will start blinking, press Hold (Up)/ Mode (Down)
button to select the desired digit (“1”, “0”, “-”).
Note: The temperature range is -40~150°F (-40~302°F),

so the meter will not function if Hi/Lo alarm setting
exceed the temperature range.

Long press MX/MI button to save the High temp. alarm
setting and “H” icon will appear on the left of the current
temp. reading.

With meter on, press MX/MI and MODE buttons
at the same time to enter Low alarm setting. A blinking “A-
on” icon will appear on the LCD.
Press MODE button to select Hi alarm “A-on” or “A-oF”.
(a) If “A-on” is selected, press MX/MI button to save and
move to step 2 for Low temp. alarm setting.
(b) If “A-oF” is selected, press MX/MI button to save and
then the meter will return to normal measurement.

After Lo alarm is set, L-40°C will appear on the LCD and
digit “1” will start blinking, press Hold (Up)/ Mode (Down)
button to select the desired digit (“1”, “0”, “-”).
Note: The temperature range is -40~150°C (-40~302°F),

so the meter will not function if Hi/Lo alarm setting
exceed the temperature range

Press MX/MI button to save the Low temp. alarm setting
and “L” icon will appear on the left of the current temp.
reading.

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

©Copyright 2012 Dwyer Instruments, Inc.

Printed in U.S.A. 8/12

FR# R6-443999-00

Advertising