Table 4-2: SPI Timing Specifications

Symbol

Description

Min

Max

Units

t

SC

Time from SSN to CLEAR

10

ns

t

CMIN

CLEAR duration

100

ns

t

SSDV

Time from SSN to Command Byte on MOSI

1

us

t

DBSH

Time to setup data before active edge

50

ns

t

DASH

Time to setup data after active edge

50

ns

t

SHZD

Time from SSN to data tri-state time

100

ns

4.5 Idle Mode

The 3D MagIC incorporates an Idle Mode to reduce power consumption, in which the circuit

automatically idles when it is not exchanging data or taking a measurement. The 3D MagIC

starts in the Idle Mode at power-up and remains in Idle Mode until a measurement is needed.

Note: The 3D MagIC starts in Idle Mode when powered on. This is different from the legacy 11096
ASIC, which required cycling through one measurement request operation to put it into Idle Mode.