5 3d magic operation – standard mode, 1 cycle count registers, 3d magic operation – PNI RM3000-F Sensor Suite User Manual

Page 29: Standard mode, Cycle count registers, Ons 5.1

Advertising
background image

PNI Sensor Corporation

Doc #1016102 r04

RM3000-f & RM2000-f Sensor Suite User Manual

Page 29

5 3D MagIC Operation

– Standard Mode

Note: This section discusses how to operate the 3D MagIC in Standard Mode. For a description of
operation in Legacy Mode, see Section 6.
Legacy Mode is intended for customers who previously used
PNI’s 11096 ASIC. The 3D MagIC operates in Standard Mode when pin #12 is held LOW.

The basic functions to be performed when operating the 3D MagIC are:

Setting the values in the Cycle Count Registers, and

Taking sensor measurements.

The user should first establish the number of cycle counts to be measured for each sensor by

writing to the Cycle Count Registers. This is followed by sending a command or series of

commands to make the sensor measurements. Assuming the user will use the same number of

cycle counts for subsequent measurements, it is not necessary to rewrite to the Cycle Count

Registers for subsequent sensor measurements.

In Standard Mode, the 3D MagIC provides two methods to take sensor measurements, which are

discussed later in this section:

Single-axis measurement (SAM), and

Multi-axis measurement (MAM).

The SAM Command Byte initiates a measurement for one sensor and sets up the 3D MagIC to

write the measured values out on the MISO line. The MAM Command Byte sets up the 3D

MagIC to measure up to 3 sensors, and a later Command Byte initiates reading the measurement

results registers for up to 3 sensors. For 2-axis or 3-axis systems, it is more efficient to operate

using the MAM Command Byte.

Note that sample code can be found in the

Appendix

.

5.1 Cycle Count Registers

Prior to sending a command to take a sensor measurement, it is necessary to write values to

the Cycle Count Registers. (The default value for the Cycle Count Registers is 512D, but

this was chosen for legacy reasons and is an inefficient value otherwise.) The Cycle Count

Registers establish the number of sensor oscillation cycles that will be counted for each

sensor in both the forward and reverse bias directions during a measurement sequence.

Increasing the cycle count value improves measurement sensitivity, but system noise limits

the useable sensitivity such that the maximum efficient cycle count value generally is around