Fatal error, Output error, Out of range error – ROTRONIC CF8-W-Disp-AL User Manual

Page 6: Memory error, Warm up state, Error code and action plan

Advertising
background image

Dokument
UMA 145

Rev
3

Page
6 (8)

Error code and action plan

Bit # Error

code

Error description

Suggested action

0

N/A

Fatal Error

Try to restart sensor by power
OFF/ON. Contact local distributor

1

2

Reserved

2

4

Algorithm Error.
Indicate wrong EEPROM
configuration.

Try to restart sensor by power
OFF/ON.
Check detailed settings and
configuration with UIP software
version 4.3 and higher.
Contact local distributor.

3

8

Output Error
Detected errors during output
signals calculation and
generation.

Check connections and loads of
outputs.
Check detailed status of outputs
with UIP software version 4.3 and
higher.

4

16

Self-Diagnostic Error.
May indicate the need of zero
calibration or sensor
replacement.

Check detailed self-diagnostic status
with UIP software version 4.3 and
higher.
Contact local distributor

5

32

Out of Range Error
Accompanies most of other
errors. Can also indicate
overload or failures of sensors
and inputs.

Resets automatically after
source of error disappearance.

Try sensor in fresh air.

Check detailed status of
measurements with UIP software
version 4.3 and higher.

See Note 1!

6

64

Memory Error
Non-fatal error during memory
operations.

Check detailed settings and
configuration with UIP software
version 4.3 and higher.

7

128

Warm Up state
Is always set after power up or
power failure. Resets after
restart sequence.

If it doesn’t disappear in half a
minute, check power stability.


Note 1

. Any probe is out of range. Occurs, for instance, during over exposure of CO2

sensor, in which case the error code will automatically reset when the measurement values
return to normal. Could also indicate the need of zero point calibration.


Remark

: If several errors are detected at the same time the different error code numbers will be

added together into one single error code!

PLEASE NOTE! The sensor accuracy is defined at continuous operation (at
least 3 weeks after installation)