Digital interface (hf65), 5 digital interface (hf65) – ROTRONIC HF65 User Manual

Page 10

Advertising
background image

E-M-HF63_65-V1_11

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

HygroFlex HF63/HF65 Humidity Temperature
Transmitters: User Guide

Document title

Page 10 of 34

© 2008; Rotronic AG E-M-HF63_65-V1_11

The D/A converters used to generate the analog output signals feature a 16-bit resolution and exhibit a small
positive offset at the bottom of the signal range as indicated below:.

Signal type

Maximum offset at range bottom

0…1 V

3 mV

0…5 V

50 mV

0…10 V

90 mV

0…20 mA

4 µA

4…20 mA

No offset


The ROTRONIC HW4 software allows changing the type of output signal to one of the following: 0…20 mA,
4…20 mA, 0…1V, 0…5V or 0…10V. Both output signals are automatically configured with the same signal
type. No calibration or adjustment is required after changing the type of output signal.

3.5 Digital interface (HF65)


a) RS-485 serial interface


The HF65 has always a RS-485 serial interface. When the number of available network ports is limited, this
interface can be used to connect together up to 64 devices in a multi-dropped arrangement. In principle, an
unlimited number of such networks can be monitored with the HW4 software, but each RS-485 multi-drop
network is limited to 64 devices. The HF65 can be used either as a slave or a master, without special
configuration. The master is automatically the device that is directly connected to a network (PC or LAN) by
means of a USB port or TCP/IP port.

RS-485 Compatibility: The communication protocol used by the HF65 and other AirChip 3000 products is not
compatible with the protocol used by the previous generation of ROTRONIC instruments. Do not connect
legacy products and AirChip 3000 products to the same RS-485 multi-drop network.

b) USB or Ethernet interface


The HF65 is also available with the following interface combinations: USB + RS-485 or Ethernet +RS-485. In
that case the protection grade of the HF6 enclosure is no longer IP65 / NEMA 4 rated.

c) Communication protocol options


The measurement data can be read without having to use the ROTRONIC HW4 software. Starting with
firmware version 1.3, the HF65 offers the following communication protocol options (ASCII) which can be
selected by connecting the HF65 to a PC running the ROTRONIC HW4 software (version 2.1.1 or higher):

o

RO-ASCII: this is the standard (default) communication protocol used by all AirChip 3000 devices and by
the HW4 software. In principle, this protocol supports all of the AirChip 3000 functions but some of the
functions require a certain amount of computations to be carried out by an external device such as a PC.

o

Custom: this communication protocol can be used to provide compatibility of the HF65 with an existing
communication system. The Custom communication protocol is limited to reading measurement data from
the HF65. Functions such as device configuration, humidity and temperature adjustment, etc. are not
supported. The Custom protocol is applicable to all AirChip 3000 devices with a digital interface and
allows RS-485 networking