RS-485 terminal block



Pins K3-1 and K3-2 can be used to power all
instruments on a RS-485 multi-drop from a single
external DC power supply with adequate mA rating.
In that case, do not use pin K6-2 and pin K6-3
(normally used to power the HF65).




6.4.3

Grounding (all models)

We generally recommend grounding the (-) side of the power supply, especially if the electronics will be
subjected to a low humidity environment (35 %RH or less).

7

Operation

7.1 Minimum load requirements for the HF63 with voltage outputs


The following requirements apply to any external device or circuit connected to the HF63 transmitter with
voltage outputs:

















Terminals

Description

K3-1: PWR

DC (+) 15…40 VDC (+)
(optional, see note below)

K3-2: GND

Ground / Power supply (-)

K3-3: D+

RS-485 Bi-directional
TX+ / RX +

K3-4: D-

RS-485 Bi-directional
TX- / RX -

RL

GND

HF63

Voltage Output

HF63 output signal

Input resistance RL

0…1V >=1kOhm

0…5V >=5kOhm

0…10V >=10kOhm