Specifications, Limited warranty – Setra System Model 470 User Manual

Page 34

Advertising
background image

34

Specifications

Pressure media:

Clean dry nonconductive noncorrosive gas

Accuracy:

accuracy

1

±

0.02% full scale at 70

°

F (21

°

C)

nonlinearity

±

0.012% full scale (End point)

hysteresis

0.010% full scale

nonrepeatability

±

0.010% full scale

thermal zero shift

2

±

0.002% FS/

°

F (0.004% FS/

°

C)

thermal span shift

2

±

0.001% FS/

°

F (0.002% FS/

°

C)

altitude resolution

1 ft. (4 ft. for 100 psi range)

stability, 24 hours

±

0.005% full scale

stability, 30 days

±

0.02% full scale

stability, 1 year

±

0.05% full scale

Output:

Bidirectional RS-232 interface

Physical:

pressure fitting

1/8” barbed male fitting

pressure port

10-32 internal thread

height

4.0”

width

3.5”

depth

5.25”

weight

2.4 pounds

Power:

5 VDC

±

1%, 70 mA; 17 mA w/200

microampere sleep mode option.

Notes:
1.

Accuracy as RSS of non-linearity, hysteresis, and non-repeatability.

2.

Unit calibrated at 70

°

F. Maximum thermal error is computed from this datum.

Information about the Model 470 Digital Pressure Transducer contained in this document is
subject to change without notice.

Advertising