Appendix B: Static calibration


1. General

- Static calibration is composed of two sub-procedures: ZERO calibration and GAIN

calibration.

- Special equipment required: A 1000-2500 gr. standard weight.

- Performing Static Calibration requires opening a password

2. Opening Password


a) While in Main Screen press the

key

"Maintenance” screen is displayed

b) Press <SET>, "Open/Close Password" screen

is displayed

c) Select (by

keys) "Open Password" and

Press <ENT>

– "Enter Password" screen is

displayed

d) Enter Password "11" (by

keys) and

press <ENT>

"Password Opened" screen is displayed

e) If the wrong password is entered' "Wrong pass"

screen will appear. Press <ESC> to return to

"Open/Close Password" screen and try again

f) Once "Password Opened" screen appears, press <ESC> twice to return to the

Maintenance screen (the password remains open!)

As long as the Password is open

– Static Calibration is possible