UEi Test Instruments DM397 User Manual

Page 12

Advertising
background image

Feature Summary

DC Voltage

NMRR

: >60dB @ 50/60 Hz

CMRR

: >120dB @ DC, 50/60 Hz, Rs = 1k

Input impedance

: 10MΩ, 30 pF nominal

(50MΩ, 100 pF nominal for 50m V
& 500mV ranges)

AC Voltage

CMRR

: >60dB @ DC to 60 Hz, Rs = 1kΩ

Input impedance

: 10MΩ, 30 pF nominal

(50MΩ, 100 pF nominal for 500m V range)

DC Current

AC Current

(AC+DC) Voltages and (AC+DC) Current

Resistance

Open Circuit Voltage: <1.3 V DC

*1.

Using relative (REL∆) mode

Conductance (5,000 counts only)

DM397-MAN

P. 11

Backlight

For clear readings in poorly lighted areas

Fast auto ranging

Meter automatically selects the best range quickly

AC+DC total RMS

Choices for AC only, AC+DC readings or AC and DC

(@40Hz to 10 kHz)

Dual display

DBm, dBV

User selectable impedance references for dBm

User selectable voltage references for dBV

Auto HOLD

Holds readings on display

Continuity/Open test

Beeper sounds

Fast analog bar graph

25 segments for view of fast events and trends

Memory locations

20

Duty cycle/pulse width

Measures the time that the signal is ON or OFF in % or

milliseconds

MIN/MAX mode

Record maximum, minimum and average values

1mS PEAK mode

Captures peaks of 1 millisecond

Close-case calibration

No internal adjustment necessary

Batter/fuse access door

Battery or fuse replaceable without voiding calibration

High-impact over molded case

Protective feature for the meter and user

Range

Resolution

Accuracy

50 mV

1 µV

0.05% + 10

500 mV

10 µV

5 V

100 µV

0.05% + 2

50 V

1 mV

500 V

10 mV

1000 V

100 mV

0.1% + 2

Range

Resolution

Accuracy

40Hz - 1kHz

4kHz - 5kHz

5kHz - 20z

20kHz - 50kHz

500 mV

10 µV

0.3%+10

1.0%+10

2.0%+20

Unspecified

5 V

100 µV

50 V

1 mV

0.3%+10

0.5%+10

0.5%+10

0.8%+20

500 V

10 mV

1000 V

100 mV

0.4%+10

0.5%+20

Unspecified

Unspecified

Range

Resolution

Accuracy

500 µA

10 nA

5 mA

100 nA

0.1% + 5

50 mA

1 µA

500 mA

10 µA

5 A

100 µA

0.3% + 10

10 A

1 mA

0.3% + 20

Range

Resolution

Accuracy

40Hz - 1kHz

1kHz - 10kHz

500 µA

10 nA

5 mA

100 nA

0.3% + 5

0.8% + 10

50 mA

1 µA

500 mA

10 µA

5 A

100 µA

0.4% + 10

Unspecified

10 A

1 mA

0.4% + 20

Function

Range

Resolution

Accuracy

40Hz - 1kHz

1kHz - 10kHz

DC mV

500 mV

100 µV

0.5% + 5

0.8% + 5

5 V

1 mV

DC V

50 V

10 mV

0.5% + 3

0.8% + 3

500 V

100 mA

1000 V

1 V

0.8% + 5

0.8% + 5

DC µA

500 µA

100 nA

5 mA

1 µA

0.5% + 3

1.0% + 5

DC mA

50 mA

10 µA

500 mA

100 µA

DC A

5 A

1 mA

0.8% + 10

Unspecified

10 A

10 mA

Range

Resolution

Accuracy

50 Ω

0.001 Ω

0.5% + 20 *

1

500 Ω

0.01 Ω

0.1% + 5 *

1

5 kΩ

0.1 Ω

50 kΩ

1 Ω

0.1% + 2

500 kΩ

10 Ω

5 MΩ

100 Ω

0.3% + 5

50 MΩ

1 kΩ

0.5% + 20

Range

Resolution

Accuracy

50 nS

0.01 nS

0.1% + 10

Advertising