UEi Test Instruments DAFM4 User Manual

Page 8

Advertising
background image

7

MEASURING AIR VOLUME (CFM OR CMM) CONT.

TIMED AVERAGE AIR VOLUME
1. Press MODE until VOL is showing in the upper right corner of

the display

2. Enter duct dimensions using the method described on page 6

3. Extend the probe to the width of the duct or more

4. Verify that you are aiming the “Air In” into the air flow. This

is the side opposite of the inch/cm markings on the probe

extension. The grooves of the probe should be perpendicular to

the air flow direction

5. Insert probe into duct access hole (3/4” hole required)

a. See “Measurement Location and Method” for instructions on

where to measure

6. Start with the probe inserted just inside the access hole

7. Press “START” to begin timed average

8. Slowly move probe into the duct until it crosses the duct and

contacts the far side

9. Press “STOP” to interrupt timed average. The cumulative timer

will stop when the DAFM4 is not capturing data

10. Move to the next access hole and repeat steps 6~9.

11. Press MIN/MAX AVG ( ) after capturing data on all access

holes to view average volume

a. AVG flashing and elapsed time will show when displaying

average

b. AVG flashing, elapsed time and MIN will show minimum

measured volume

c. AVG flashing, elapsed time, and MAX will show maximum

measured volume

12. Press and hold EXIT to end test

MEASUREMENT LOCATION AND METHOD

There are recommended methods to traverse a duct in order to collect

sufficient data for an accurate average. Using a timed average increases

the number of data points for a much more accurate result. It is still

recommended to use the approach taken by the Log-Tchebycheff Rule

for access hole spacing, but the number of data points is not needed. By

utilizing a timed average you will capture these locations, plus data from

the entire width of the duct being measured.

Advertising