F75.9, Single input: (top view), F76.1 – UEi Test Instruments DTK2KIT User Manual

Page 2

Advertising
background image

3

LCd dISPLAY

1. Primary LCD: Display T1 (1st thermocouple reading) or T1 - T2
2. Secondary LCD: Display T2

4

OPERATION

POWER ON/OFF
Note:

Make sure you have plugged in the thermocouple probe before

turning on.
Press and hold PWR button to turn on the meter. While the meter is

on, press PWR button to turn off the meter.

SLEEP MODE (AUTO POWER OFF)
The meter will turn off automatically after 15 minutes if no buttons

are pressed. To disable auto power off function, when the meter

is off, press and hold “HOLD” button then press “PWR” button

to turn on the meter, the primary LCD shows the following. Sleep

mode is now disabled.

TAKING MEASUREMENT
Connect thermocouple. If probe is open or not connected, the LCD

will show ”---”.

DIFFERENTIAL (T1 - T2)
Press "MODE" to display differential. Press "MODE" again to

return to normal display.

• MAX / MIN
Press MAX / MIN button to change operation mode from live

data -> MAX -> MIN -> back to live data. Corresponding value will

display on the LCD. These values will reset at power on or when

performing a reset on MAX/MIN.

Reset MAX/MIN: Press Max/Min button for 2 seconds to reset

MAX/MIN analysis value. The MAX/MIN value is updated to

current reading.

• DATA HOLD
Freeze display of basic measurements or disable. Press HOLD

button again to go back to the temperature reading.

CHANGE UNIT
Press UNIT button to change unit ˚C or ˚F. The meter default

temperature unit is ˚F. The meter will default to last selected scale.

1

INTROdUCTION

Thank you for purchasing the UEi DTK2 dual input K type

differential thermometer. This thermometer features a built-in

microprocessor and high resolution analog to digital converter

chip. The thermocouple probe responds very quickly to a wide

range of temperature measurements. The industry standard mini

thermocouple jack accepts a wide selection of probe types.

FEATURES
• Dual input with differential (T1 - T2)

• Min / Max capture (T1 & T2)
• Auto power off with disable feature (15 minutes)

• Large LCD screen

• Complete with bead wire temperature probes and 9V battery

• ˚C / ˚F select function

• Magnetic mount

INCLUDES
• Meter

• 9V battery

• Operation manual

• ATT29A K-type bead probe (2)

• Pouch

2

KEYPAd (CONTROLS)

T1

-

T2

HOLD MAX MIN

K

K

2

1

MODE

SET

PWR

MAX

MIN

UNIT

HOLD

PRESS “UNIT” TO SELECT ˚C/˚F.

Freeze display

Press “MODE” to change operation mode T1 or T1-T2.

The corresponding reading will display on the primary

LCD. The secondary LCD will always display T2.

Press and hold “PWR” to turn on the meter or

press it to turn off the meter. The meter will turn off

automatically after 15 minutes if you don’t press any

key. Press “HOLD” and “PWR” buttons to disable

Auto sleep mode.

Press “MAX/MIN” to change operation mode

from live data -> Max.-> Min.-> back to live date. The

corresponding value will display on the LCD. The value

is calculated since powered on or reset.

Press and hold to enter ice-point calibration. (Please

refer to section

5

to ensure correct calibration.)

Press UNIT

K

K

K

K

Single Input: (Top View)

K

K

T2

K

K

MIN

K

K

MAX

K

K

HOLD

K

K

5

AUTOMATIC ICE-POINT CALIBRATION

The DTK2 has an automatic ice-point calibration function. Please

follow these steps to calibrate your meter to the thermocouple

probes.

• ITEMS NEEDED:

1. DTK2 Thermometer

2. Two k-type probes

3. Crushed ice/water mixture

(distilled water recommended if available)

• CALIBRATION PROCEDURE
1. Connect both probes to the DTK2
2. Place probe tips in the crushed ice/water mixture
3. Power on the DTK2
4. Press and hold SET for five seconds.
a. The meter will enter ice-point calibration and 0.0°C will

flash on both T1 and T2 readings
5. Stir the ice bath with the probes submerged for one minute
6. Continue stirring the ice/water mixture until the readings

return to normal measurement (flashing will stop)

The ice-point calibration is now complete.

DTK2

75.9

°

f

75.9

°

f

T1
K

T1
K

76.1

°

f

76.1

°

f

T2
K

T2
K

Differential Thermometer

WARNING!

Always connect thermocouple before turning the

unit on for most accurate readings.

WARNING!

Make sure to plug sensor with correct polarity.

Do not be alarmed if erratic readings appear on display prior

to normal mode.

MODE

K

K

K

K