Proposed presentation contexts, 2 proposed presentation contexts – AGFA HE/001132 User Manual

Page 15

Advertising
background image

HE/001132

Page 15 of 92

Document No.001132, Revision 2.0

NodeID Livelink : 14140190

Agfa HealthCare

14 June, 2007

Table 2.2-7: Association Reject Reasons

Result

Response

Reason/Diag

The Dicom library refuses the association

SCU Device not known

The Dicom library refuses the association

Maximum number of association
processing threads exceeded

The Dicom library refuses the association

A799

Dicom Library is not allowed to accept
C-STORE commands from this device
and therefore refuses the association.

The Dicom library refuses the association

A0102

SOP Class is not found back in the
configuration

The Dicom library refuses the association

A702

The Dicom library is unable to create the
Dicom media file due to the fact that the
disk is full

The Dicom library refuses the association

A703

The Dicom library is unable to create the
Dicom media file due to resource
problems other than ‘disk is full’.

The Dicom library refuses the association

A703

The Dicom library is unable to acquire
the complete C-STORE request due to
network problems.

The Dicom library refuses the association

A701

The Dicom library is unable to acquire
the complete C-STORE request due to
network problems (time-out while
reading data from socket)

2

.

2.2.1.3.1.2

Proposed Presentation Contexts

The DICOM Store AE is capable of proposing the Presentation Contexts shown in the following table:

2

The time-out value is hard-coded in the Dicom library as being 60 seconds.