Table 2.2-29: Supported N-SET Attributes for a Basic Grayscale Image Box

Attribute Name

Tag

Value

Image Position

(2020,0010)

1

Polarity

(2020,0020)

NORMAL, REVERSE

Basic Grayscale Image Sequence (2020,0110)
>Samples Per Pixel

(0028,0002)

1

>Photometric Interpretation

(0028,0004)

MONOCHROME2

>Rows

(0028,0010)

larger than 0

>Columns

(0028,0011)

larger than 0

>Bits Allocated

(0028,0100)

8 – 16

>Bits Stored

(0028,0101)

8 – 16

>High Bit

(0028,0102)

7 to 15

>Pixel Representation

(0028,0103)

0

>Pixel Data

(7FE0,0010)

A data stream of the pixel samples that comprise the Image.

"Workstation" (SCU) will process the N-SET confirmation and response Status codes. The status codes
listed below in Table 2.2-30are recognized:

Table 2.2-30: N-SET Status Codes

Code

Status

Meaning

0000

Success

Normally returned by the SCP AE. Image successfully stored in Image Box

B604

Image size is larger then Image Box size, the image has been magnified.

B605

Warning

Requested Min Density or Max Density outside of printer’s operating range. The printer will
use its respective minimum or maximum density value instead.

C603

Image size is larger then Image box size

C605

Error

Insufficient memory in printer to store the image.

*

Other Status
codes

Other Dicom error codes result in the failure of the job. Other warnings are not communicated
to the user.

2.2.2.3.1.3.4

Printer SOP Class (1.2.840.10008.5.1.1.16)

The Printer IOD is an abstraction of the hard copy printer and is the basic Information Entity to monitor the
status of the printer.
The Printer SOP Instance is created by the SCP during start-up of the hard copy printer and has a well-
known SOP Instance UID.

The Printer SOP Class is used to monitor the status of the printer.

"Workstation" (SCU) will accept the following DIMSE services:

·

N-EVENT-REPORT

"Workstation" (SCU) can send the following DIMSE services:

·

N-GET

N-EVENT-REPORT is used to report the changes of the printer status in an asynchronous way.
The SCP uses the N-EVENT-REPORT to inform "Workstation" about each execution change. "Workstation"
will return the confirmation of the N-EVENT-REPORT operation.