18 Installation

Figure 2-1. Power Module Line Fuse and Switches

Figure 2-2. Power Module Configuration Switch

Table 2-1. Settings for Power Module Configuration Switch

Required Switch Setting

Desired Function

1

2

3

4

5

6

7*

8*

RI disabled

0

0

1

1

RI unlatched (live)

0

1

1

1

RI latching †

1

1

1

1

Display disabled

0

1

1

Display enabled †

I

1

1

Inhibit calibration

0

0

1

1

Factory calibration

0

1

1

1

Normal calibration †

1

1

1

1

Power on state is *RCL 0

0

1

1

Power on state is *RST†

1

1

1

†= factory default setting "0" Off or open "1" On or closed.
*Positions 7 and 8 are for service functions. For normal operation, they must be set to 1.