36 Calibration

Equipment Required

The equipment listed in Table A-1, or equivalent, is required for calibration.

Table A-1. Equipment Required for Calibration

Equipment

Characteristics

Recommended Model

Voltmeter

D-c accuracy 0.005%, 6 digits; resolution 1

µ

V

Agilent 3458A

Shunt

Agilent 66101A,
66102A

0.01

, 100 A, 100 W, 0.04% @ 100 W, 0.01% @ 1W,

power coeff. 0.0004%/watt in air

Guildline 9230/100

Agilent 66103A,
66104A, 66105A,
66106A

0.1

, 15 A, 25 W, 0.04% @ 25 W, 0.01% @ 1W, power

coeff. 0.002%/watt in air

Guildline 9230/15

GPIB Controller or
Keyboard

HP Vectra (or IBM compatible) with GPIB Interface or Agilent BASIC series Agilent 66001A
MPS

Calibrating Voltage

Calibration may cause dangerous voltages to be present at the power module output.



The following procedure calibrates the output voltage and the overvoltage protection (OVP). During voltage calibration,
you will enter two readings from the external DVM. You must calibrate the voltage before calibrating the OVP. You do
not have to enter any external readings when calibrating the OVP.

Table A-2. Voltage Calibration Procedure

Keyboard Entry

Controller Entry

Function

Connect the equipment as shown in Figure A-lA

Output Voltage Calibration

*RST

Reset the power module

(Before proceeding, set the module output, via the keyboard or the GPIB, for 1.0 V at 0.5 A)

OUTP ON

Enable the output

CAL:STAT 1,66102

Using your password, enable the calibration
mode

CAL:VOLT:LEV MIN

Select the lower calibration point

CAL: VOLT: DATA <NRf>

Enter the digital voltmeter reading

CAL:VOLT:LEV MAX

Select the upper calibration point

CAL:VOLT:DATA <NRf>

Enter the digital voltmeter reading


Note:

The following step overwrites the existing voltage calibration constants stored in nonvolatile memory.

CAL:SAV

Save the new calibration constants

Overvoltage Protection Calibration
If the optional output relay is installed, either disconnect it or set it to the OFF (open) state.

CAL:VOLT:PROT

Perform OVP calibration


Note:

It takes a few seconds for the power module to make the measurement. When completed, the new constant is

automatically stored in nonvolatile memory.

See Figure A-2 for program listing.