Newlife, Intensity – AirSep MN139-1 User Manual

Page 67

Advertising
background image

NewLife

®

Intensity

F

AirSep

®

Corporation

IV. 6. technical characteristics

Dimensions: D x W x H: 368 x 419 x 699 mm.

Weight: 25.0 kg.

Noise level: ± 52 dBA

Flow values:

0 - 8 l/min.

average oxygen content:

• at 2 l/min. 92 %.

• at 7 l/min. 91 %.

• at 8 l/min. 90%.

(values at 21 °C and at an atmospheric pressure of 1013 mbar). Max. recommended flow:

8 l/min. The maximum outlet pressure is 138 kPa. It is not recommended to use the NewLife

®

Intensity at flow values less than

2 l/min. unless utilized with the Pediatric Option Procedure.

electrical power supply:

• 230 V - 50 hz europe / 115 V – 60 hz / 230 V – 60 hz (other countries,

depending on version)

• Use the cable provided, and check that the electrical characteristics of the mains socket used

match those indicated on the manufacturer’s plate on the rear panel of the machine.

• Mean power rating: 410 watts

Class II

Type B

2.0 amps (230 V).

4.0 amps (120 V).

Filters:

At the rear of the device: a dust filter.

Before the oxygen outlet: a product filter < 2.0 µm.

air circulation:

A fan cools the compressor compartment.

environmental limit conditions:

The performances of the device (especially the oxygen concentration) are quoted at 21 °C and

1013 mbar. They may change with temperature and altitude.

- The device should preferably be stored and transported in the vertical position.

- The device must only be used in the vertical position.

- Ambient temperature of between 5°C and 40 °C (operation).

- Storage temperature range from -20 °C to 60 °C.

- Relative humidity of between 0% and 95% (operation and storage).

- IPX1: Protected against dripping water (complies with the EN 60601-1 standard)

7-Gb

GB

Advertising