AirSep MN138-1 H User Manual

Page 34

Advertising
background image

28

MN138-1 Rev H

02/14

5.3 Troubleshooting Chart

Problem

Probable Cause

Solution


Unit does not run.
I/0 power switch in “I” position.

•

No power to unit.

•

Unit circuit breaker tripped/faulty

•

Faulty electrical connections.

•

Faulty circuit board.

•

Faulty I/0 power switch.

•

Check wall outlet for power.

•

Reset or replace circuit breaker.

•

Check electrical connections.

•

Replace circuit board.

•

Replace I/0 power switch.


Compressor runs with intermittent
alarm. Oxygen concentration at max
flow rate is within specifications.



• Faulty pressure transducer.



• Replace circuit board.

Compressor runs and shuts down
periodically. Only start up alarm
activates


• Faulty oxygen monitor circuit board.


• Replace oxygen monitor circuit board.


Compressor runs and shuts down
periodically with 5-beep intermittent
high temperature alarm.

•

Restricted air flow.

•

Unit overheating due to improper
location. Compressor thermally cut off
due to excessive heat. NOTE: It will
not restart until it cools down.

•

Faulty cabinet fan.

•

Temperature probe wire harness faulty
or disconnected from circuit board.

•

Remove obstruction.

•

Locate unit away from heating source,
providing adequate ventilation on all
sides.

•

Do not operate more than one unit
side by side.

•

Replace cabinet fan.

•

Replace or reconnect temperature
probe.


Compressor runs with 3-beep
intermittent low pressure alarm
and low oxygen concentration.

 

•

Leak

 

•

Reduced air intake

•

Weak compressor

•

Faulty circuit board.

•

Leak test and repair leak.

•

Check compressor intake path for
obstruction. Clean/remove
obstruction.

•

Replace compressor.

•

Replace circuit board

.



Compressor does not start. I/0
power switch in “I” position,
intermittent 3-beep low pressure
alarm, and cabinet fan turns.



•

Extreme cold start.

•

Faulty electrical connection for
compressor.

•

Faulty capacitor.

•

Faulty circuit board.


•

Allow unit to reach room
temperature.

•

Check electrical connections for
compressor.

•

Replace capacitor.

•

Replace circuit board

4-beep high pressure alarm is
activated and both compressor
and valves do not operate.


•

Restriction in exhaust muffler.

•

Contaminated sieve beds.

•

Faulty solenoid valve.

•

Faulty circuit board.


•

Replace exhaust muffler.

•

Replace sieve beds.

•

Replace solenoid valve manifold.

•

Replace circuit board.