Select an Analysis Mode

5. Press

the

SELECT

key to blink a symbol, it will

blink. Select an analysis mode using the

↓

or

↑

key.

If standard mode is used, select

5td

.

If automatic mode is used, select

U-a

.

If manual mode is used, select

U-m

.

Set ACCURACY for the Standard Mode

6. Select ACCURACY with the

SELECT

key.

Select MID. with the

↓

or

↑

key.

When

ACCURACY

is selected, HI, MID. or LO blinks.

Set the Heating Pattern

7. Select drying temperature 1 with the

SELECT

key. Select step drying

of the heating

pattern with the

PROGRAM

key.

Set Drying Temperature 1

8. Set 120°C to drying temperature 1 with

the

↓

or

↑

key.


Set the Time

9. Select time with the

SELECT

key.

Set 5.0 minutes with the

↓

or

↑

key.


Set Drying Temperature 2

10. Select drying Temperature 2 with

the

SELECT

key.

Set 160°C with the

↓

or

↑

key.



Set the Preset Termination Value for Automatic Mode

11. Select a preset termination value with the

SELECT

key. Select 0.05 [%/min] with

the

↓

or

↑

key.


Set Measurement Unit

12. Select measurement unit with the

SELECT

key.

Select moisture content (based on a wet
sample ) with the

↓

or

↑

key.


ACCURACY

MID.

Standard

mode

"0."0"5

%/min

Automatic

mode

"U"-"a

0.000

g

0

"%

"M"O"I"S"T
"/"W

"5."0 min

2

"1"6"0°C

"1"0"5°C

1

"1"2"0°C

Standard mode
Manual mode

Automatic mode
Manual mode

ACCURACY

Preset termination value

Analysis

Measurement unit

Drying temperature 2

Drying temperature 1

Step drying

Time