Technical specifications – Vaisala HM34 User Manual
Page 7
5. TECHNICAL SPECIFICATIONS
Measurement range
humidity
0...100 %RH
temperature
-20...+60
°
C or
-4...+140
°
F
Accuracy (at +20
°
C):
(including nonlinearity and hysteresis)
relative humidity
against factory references
+1 %RH
against field references
+2 %RH (0...90 %RH)
+3 %RH (90...100 %RH)
temperature
+0.3
°
C (+0.5
°
F)
NOTE:
When the HM34 is used in an electromagnetic field of 3 V/m,
with the frequency of 200 - 220 Mhz, the temperature
deviation is max. ±3 °C.
Temperature dependence:
+0.04 %RH/
°
C
+0.02
°
C/
°
C
Resolution:
0.1 %RH
0.1 °C (0.1 °F)
Response time for
humidity (at +20 °C):
15 s
Display:
3
1
/
2
-digit LCD
HOLD function:
pushbutton hold
for displayed value
Supply voltage:
9 V battery (IEC 6F22)
Battery operating time:
50 h (typ.)
Automatic power-off:
after 3 minutes, unless
HOLD is activated
5