Press enter repeatedly until the following appears, Two- point humidity calibration, Feeding the offset correction – Vaisala HM44 User Manual

Page 38: Feeding the gain correction

Advertising
background image

User's Guide _______________________________________________________________________

_________________________________________________________________________________

36

After a few seconds, the text changes automatically to show the following:

set

°C

Press ENTER repeatedly until the following appears:

set

This indicates that no calibration has been selected. In addition to this, there are five
calibration types available. Select the desired type with buttons

▲ or ▼. All selections are

acknowledged with ENTER. In the table below, you can find a summary of all the calibration
options.

RH

set

T

Default calibration restores the factory settings of the
humidity and temperature calibrations.

RH

set

One point humidity calibration; select this for performing
a humidity calibration at one point (not instructed).

RH

set

Two point humidity calibration; select this for performing
a humidity calibration at one point (see section

Two-

point humidity calibration

on page 37).

RH

set

Entering the offset correction manually; corresponds to
the mode explained in
Section

Feeding the offset correction

on page 8.

RH

set

Entering the gain correction manually; corresponds to the
mode explained in
Section

Feeding the gain correction

on page 8.