Two-Point Adjustment Using a
Calibrator

Note that the difference between the two reference humidities must be
at least 50 %. MI70 indicator is used now only as a terminal for
visualizing and setting the transmitter's RH reading.

Follow steps 1 to 7 on page 42 and continue as follows:

8.

Remove the filter from the transmitter's probe and insert the
probe into the lower humidity reference condition.

9.

Press

ADJUST

to continue adjusting.

10. Choose 2-point adjustment from the MI70 adjustment menu,

press

SELECT

.

11. Press

READY

when the reading has stabilized in the first reference

condition (may take 30 minutes or more). You can follow the
stabilization from the

GRAPH

display.

12. Enter the correct reference value in the first condition with the

arrow buttons. Press

OK

.

13. Remove the probe from the first reference conditions and insert

the probe into the higher humidity reference condition.

14. Press

READY

when the reading has stabilized in the second

reference condition (may take 30 minutes or more). You can
follow the stabilization from the

GRAPH

display.

15. Enter the correct reference value in the second condition with

the arrow buttons. Press

OK

.

16. Confirm the adjustment by pressing

YES

(by pressing

NO

you

return to adjustment mode display and no changes are made). If
the difference between the two reference conditions is less than
50 %RH, adjustment can not be done.

17. The adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

18. Turn off the MI70 and detach the connection cable.