Vaisala HMDW110 User Manual

Page 37

Advertising
background image

Syntax

Description

asel ?<cr>

Show currently set analog output
parameters and scaling.

asel<cr>

Show currently set analog output
parameters and scaling, prompt to enter new
scaling values.

asel [ch1 ch2]<cr>

Set new output parameters for both
channels, prompt to enter new scaling
values. Selectable parameters for ch1 and
ch2

are:

t

= temperature

rh

= relative humidity

td

= dewpoint

h

= enthalpy

tw

= wet bulb temperature

asel [ch1 ch2 

[ch1lo ch1hi ch2lo ch2hi]]<cr>

Set analog output parameters and scaling for
both channels without prompting.

ch1

= Output parameter for channel 1.

Parameters as above.

ch2

= Output parameter for channel 2.

Parameters as above.

ch1lo

= Low limit for channel 1 scaling.

ch1hi

= High limit for channel 1 scaling.

ch2lo

= Low limit for channel 2 scaling.

ch2hi

=High limit for channel 2 scaling.

Example (show current analog output parameters and scaling):

asel ?
Ch1 RH

lo

:

0.00 %RH

Ch1 RH

hi

:

100.00 %RH

Ch2 T

lo

:

-40.00 'C

Ch2 T

hi

:

60.00 'C

Table 27 Asel Command

35

4 Service Port

Advertising