Dip switches, Figure 19, Dip switches on component board – Vaisala HMDW80 User Manual
Page 32
Advertising

User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M211510EN-C
DIP Switches
The DIP switches on the component board control the humidity output
parameter and scaling, and the units that are shown on the display.
The switches are present on the following models:
- HMD82(D) and HMD83(D)
- HMW88(D) and HMW89(D)
- HMS82 and HMS83
Figure 19 below shows the default position of the DIP switches. The
default settings are:
- Metric units shown on display.
- Relative humidity (RH) selected as humidity output parameter.
1302-053
Figure 19
DIP Switches on Component Board
1
2
3
4
Metric
RH
Non-metric
Td
h
Tw
ON