Set/show calibration date, Set/show calibration information – Vaisala HMT130 User Manual

Page 34

Advertising
background image

User's Guide _______________________________________________________________________

32 ___________________________________________________________________ M211280EN-B

NOTE

Single-parameter transmitters HMT130H and HMT130T have only one

quantity (T for HMT130T, and RH, TD, TDF, A, X, H, TW, PWS, PW

for HMT130H.)

Set/Show Calibration Date

Use the CDATE command to show or set the calibration date field.

(Calibration refers to the calibration of the HMT130's analog outputs.)

CDATE [date]<cr>

where
date = Calibration date in format yyyy-mm-dd

Examples:

>cdate

Calibrated : "NOT SET"

>

>cdate 2009-06-22

Calibrated : "2009-06-22"

>

Set/Show Calibration Information

Use the CTEXT command to show or set the calibration info field.

(Calibration refers to the calibration of the HMT130's analog outputs.)

CTEXT [location]<cr>

where
location = Freely defined text string typically indicating the

calibration location

Examples:

>ctext

Calib. info: "VAISALA HELSINKI"

>

>ctext "Calibration lab 2"

Calib. info: "Calibration lab 2"

>