List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
In the tables below, the bold text in the brackets is the default setting.
Table 6
Measurement Output
Command
Description
R
Start continuous outputting
S
Stop continuous outputting
INTV [
0 … 255 S/MIN/H]
Set continuous output interval for RUN
mode
SEND [
0 … 99]
Output reading once
SMODE [
STOP/RUN/POLL]
Set serial interface mode
SERI [baud p d s]
Serial line settings (default: 4800 E 7 1)
baud: 300…19200
ADDR [
0 … 99]
Set transmitter address for use in POLL
mode
OPEN [0 … 99]
Temporarily open transmitter from POLL
mode to receive commands
CLOSE
Set transmitter in POLL mode
Table 7
Output Formatting
Command
Description
FORM
Set serial output format
TIME
Set time
DATE
Set date
FTIME [ON/
OFF]
Add time to SEND and R outputs
FDATE [ON/
OFF]
Add date to SEND and R outputs
UNIT
Select metric or nonmetric output units
FST [ON/
OFF]
Output state of optional probe heating and
chemical purge (with SEND and R
commands)