List of Serial Commands

All commands can be issued either in uppercase or lowercase. In the

command examples, the keyboard input by the user is in bold type.

The notation <cr> refers to pressing the carriage return (Enter) key on

your computer keyboard. Enter a <cr> to clear the command buffer

before starting to enter commands.

In the tables below, the bold text in the brackets is the default setting.

Table 6

Measurement Output

Command

Description

R

Start continuous outputting

S

Stop continuous outputting

INTV [

0 … 255 S/MIN/H]

Set continuous output interval for RUN

mode

SEND [

0 … 99]

Output reading once

SMODE [

STOP/RUN/POLL]

Set serial interface mode

SERI [baud p d s]

Serial line settings (default: 4800 E 7 1)

baud: 300…19200

ADDR [

0 … 99]

Set transmitter address for use in POLL

mode

OPEN [0 … 99]

Temporarily open transmitter from POLL

mode to receive commands

CLOSE

Set transmitter in POLL mode


Table 7

Output Formatting

Command

Description

FORM

Set serial output format

TIME

Set time

DATE

Set date

FTIME [ON/

OFF]

Add time to SEND and R outputs

FDATE [ON/

OFF]

Add date to SEND and R outputs

UNIT

Select metric or nonmetric output units

FST [ON/

OFF]

Output state of optional probe heating and

chemical purge (with SEND and R

commands)