Ashcroft 2274 Digital Industrial Gauge User Manual

Page 14

Advertising
background image

– 14 –

Step 2: Press the

(up arrow key) or

(down arrow

key) until the word UPDATE appears.

Step 3: Press ENTER.

Step 4: Press the

(up arrow key) or

(down arrow

key) to select an update rate.

Step 5: Press ENTER to finalize the selection.

DAMP or dampening: with five different options, this mode

allows for taking process pressure readings and averaging

them. This option is particularly useful to stabilize minor

process fluctuations. The options are NONE*, AVG 8, AVG

6, AVG 4, AVG 2.

Step 1: Press the MENU key until the word dAMP appears.

Step 2: Press ENTER.

Step 3: Press the

(up arrow key) or

(down arrow

key) to select a dampening option.

Step 4: Press the ENTER key to finalize your Damp option.

(This Menu item is only seen on units with the switch option)

SWSET: Allows for setting switch setpoints. The gauge is

offered with (1) or (2) SPDT switches. If (1) SPDT switch

is ordered the menu option is SW1. If (2) SPDT switches

are ordered, the MENU options are SW1 and SW2.

Step 1: Press the MENU key.

Step 2: Press the

(up arrow key) or

(down arrow

key) on the keypad to select the switch to be set. (If two

switches are present.)

Step 3: Press ENTER. The top line of the gauge display will

indicate pressure at 60% of the full-scale gauge range* or

the most recent switch setpoint. The middle line of the dis-

play will indicate a bar graph that displays the pressure

MENU FUNCTIONS

*Indicates default.

I&M008-10109-DIG Rev F 07-12_layout 10/14/13 9:45 AM Page 14