5 long lead lengths – Campbell Scientific 03001-L R.M. Young Wind Sentry Set User Manual
Page 15

R.M. Young Wind Sentry
7: If (X<=>F) (P89)
1: 3
X Loc [ WindDir ]
2: 3
>=
3: 360
F
4: 30
Then
Do
8: Z=F x 10^n (P30)
1: 0
F
2: 0
n, Exponent of 10
3: 3
Z Loc [ WindDir ]
9: End (P95)
10: If (X<=>F) (P89)
1: 3
X Loc [ WindDir ]
2: 4
<
3: 0
F
4: 3
Then
Do
11: Z=F x 10^n (P30)
1: 0
F
2: 0
n, Exponent of 10
3: 3
Z Loc [ WindDir ]
12: End (P95)
13: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
14: Set Active Storage Area (P80)^15464
1: 1
Final Storage Area 1
2: 101
Array
ID
15: Real Time (P77)^6687
1: 1220
Year,Day,Hour/Minute (midnight = 2400)
16: Wind Vector (P69)^28601
1: 1
Reps
2: 0
Samples
per
Sub-Interval
3: 0
S, theta(1), sigma(theta(1)) with polar sensor
4: 2
Wind Speed/East Loc [ WS_ms ]
5: 3
Wind Direction/North Loc [ WindDir ]
5.5 Long Lead Lengths
When sensor lead length exceeds 100 feet, the settling time allowed for the
measurement of the vane should be increased to 20 milliseconds.
11